Centrifuges
The centrifuge is a commonly used tool in laboratory research. It uses centrifugal
force to separate substances in liquid or solid media according to particle size and
density differences. Centrifugation may present two serious hazards: mechanical failure
and dispersion of aerosols. Therefore, training on how to use the centrifuge properly
and safely is essential for all new employees as part of Lab-Specific Training.
Safe Procedures for Centrifugation
Before Centrifugation
- Train each operator on proper operating procedures, review the user manual
- Use only rotors compatible with the centrifuge. Check the expiration date for ultracentrifuge
rotors
- Check tubes, bottles, and rotors for cracks and deformities before each use
- Make sure that the rotor, tubes, and spindle are dry and clean
- Examine O-rings and replace if worn, cracked, or missing
- Never overfill centrifuge tubes (don't exceed ¾ full)
- Always cap tubes before centrifugation
- Always balance buckets, tubes, and rotors properly
- Check that the rotor is seated on the drive correctly, close the lid on the centrifuge,
and secure it
- When using swinging bucket rotors, make sure that all buckets are hooked correctly
and move freely
During Centrifugation
- Keep the lid closed at all times during operation. Never open a centrifuge until the
rotor has stopped
- Do not exceed safe rotor speed
- The operator should not leave the centrifuge until full operating speed is attained
and the machine appears to be running safely without vibration
- Stop the centrifuge immediately if an unusual condition (noise or vibration) begins
and check load balances
After centrifugation
- Allow the centrifuge to come to a complete stop before opening
- Wear gloves to remove rotor and samples
- Check inside of centrifuge for possible spills and leaks, clean centrifuge and rotor
thoroughly if necessary
- Wash hands after removing gloves
Centrifuging Infectious Materials or Human Samples
Follow the safety procedures above, plus:
- Place a biohazard label on the centrifuge
- Always wear gloves when handling tubes or rotors
- Avoid the use of celluloid tubes with biohazards. If celluloid tubes must be used,
an appropriate disinfectant must be used to decontaminate them
- Always use sealed safety cups, safety buckets, or sealed rotors with O-ring as secondary
containment if available
- Fill centrifuge tubes, load into rotors, remove from rotors, and open tubes within
a biological safety cabinet if biological safety cabinet is available
- Wipe exterior of tubes or bottles with disinfectant prior to loading into rotor or
bucket. Seal rotor or bucket, remove outer gloves, and transport to the centrifuge
- Always wait at least 10 minutes after the run to allow aerosols to settle before opening
the centrifuge. Check for possible spills or leaks
- Decontaminate centrifuge interior, safety cups or buckets, and rotors if tube breakage
occurs
- Include centrifugation procedure and decontamination plan in lab SOPs
Centrifuge Maintenance
Moisture, chemicals, strong cleaning agents, and other substances can promote corrosion
of centrifuge parts and cause centrifuge failure. The following are general maintenance
recommendations:
- Follow manufacturer instructions for maintenance and cleaning
- Keep the centrifuge clean and dry
- Cleanup all non-infectious spills immediately
- Decontaminate the rotor after use with biological or radioactive materials. Use 10%
bleach for 30 minutes followed by 70% ethanol and let air dry to clean rotors and
cups
- Never clean rotors and associated parts with abrasive wire brushes
- Store the rotor upside down in a dry place, with lids or plugs removed to prevent
condensation
- Remove adapters after use and inspect for corrosion
- Inspect rotor regularly. Remove rotors from use that show any sign of defect, and
report it to a manufacturer's representative for inspection
Maintaining a Log Book
To avoid rotor failure, keep a log book for high speed rotors recording the length
of time and speed for each use. Track and discard rotors according to the manufacturer's
recommended schedule.
Centrifuge Emergency Procedures
The following events are considered an emergency:
- If there is a spill in the centrifuge
- If centrifuge malfunctions
- If there is a rotor failure
- If there is a tube breakage
Emergency Procedures
- Turn centrifuge off immediately and close the centrifuge lid
- Notify others and evacuate the lab
- Close the lab door
- Post a biohazard spill sign at the lab door
- Leave for 30-minutes to reduce the risk of aerosols
- For spill cleanup, the operator must wear proper gloves, remove debris, clean and
disinfect centrifuge interior, rotors, safety cups or buckets following the manufacturer's
instructions
- Place any contaminated protective clothing, gloves, and all cleanup materials in a
biohazard bag
- Wash hands and any exposed skin surfaces with soap and water
- Seek medical attention as necessary
- Report incidents to P.I. or lab supervisor and the Office of Safety. Call Lab & Research
Safety Officer 701.777.3341 for assistance